Liquid-Lens Focus Calibration Tool

Builds a z → diopter lookup table for an Optotune liquid lens, using a multi-camera Basler rig to measure the true z of static AprilTag targets.

Hardware

Device Role
6× Basler cameras Triangulate AprilTag z position
XIMEA CB160CG-LX-X8G3 Measure focus sharpness through the liquid lens
Optotune liquid lens Swept in focal-power (diopter) mode with temperature compensation

Calibration data: /home/nfc/braid-configs/calibration_charuco.xml

To generate an AprilTag sheet: shiqiliu-67.github.io/apriltag-generator

Installation

uv sync

The optotune-lens package is sourced from ../optotune-lens (local path). Basler Pylon SDK and XIMEA xiAPI runtime must be installed system-wide.

Getting Started

The liquid lens focuses at a different real-world distance for every diopter value you command it to. This tool builds the lookup table between the two: it places an AprilTag at a known distance (measured by triangulating it with the six Basler cameras), sweeps the lens through its diopter range while watching a separate XIMEA camera behind the lens, and records the diopter at which that tag is sharpest. Do this at enough distances and a z → diopter curve is fit at the end — that curve is what a real application uses at runtime to autofocus.

To run a calibration session:

  1. Place an AprilTag (family 36h11) somewhere in view of the Basler rig.
  2. Start the tool: uv run lens-calibrate.
  3. Press SPACE in the preview window — it triangulates the tag, sweeps the lens, and records the best-focus diopter for that position.
  4. Move the tag to a new distance and press SPACE again. Repeat for ~10–15 positions spanning the range you care about.
  5. Press Q to quit. This fits the z → diopter curve and writes a timestamped CSV of every raw measurement.

See Procedure below for the full session walkthrough, and Output for what's in the CSV.

Usage

uv run lens-calibrate [options]
Flag Default Description
--calibration /home/nfc/braid-configs/calibration_charuco.xml Braid multi-camera calibration XML. If the default doesn't exist and no --calibration is given, the tool errors out rather than silently proceeding
--port auto Optotune lens controller serial port. Auto-detected by probing both /dev/optotune_icc1c and /dev/optotune_ld (each tried as both ICC-1C and Driver 4); set explicitly to restrict detection to one port
--exposure 10000 XIMEA exposure in µs
--coarse-steps 20 Steps in the full-range diopter sweep
--fine-steps 40 Steps in the narrow fine sweep per tag
--fine-repeats 1 Repeats per fine-sweep direction (hi→lo / lo→hi), for hysteresis stats
--settle-ms 25 Wait after each diopter change (ms). Minimum 25 ms — the lens driver is open-loop and needs that long to physically settle; lower values are rejected. Default is the fastest safe value
--z-thresh 0.02 Max z-spread (m) to treat tags as coplanar
--tag-family 36h11 AprilTag family (36h11, 25h9, …)
--debug off Save focus-curve plots and ROI crops to ./debug/ after each sweep

Procedure

Target setup

Use one or more AprilTag (family 36h11) markers. Two modes are auto-detected based on the spread of triangulated z values:

Single-height mode — all tags within --z-thresh of each other: - Z values are fused (weighted by number of cameras per tag). - One (z, diopter) data point is recorded per press of SPACE. - Move the target to a new height and repeat (~10–15 positions for a good fit).

Multi-height mode — tags span more than --z-thresh in z: - Each tag is triangulated and focused independently. - One (z, diopter) data point per tag per press of SPACE. - Useful for a stacked multi-plane target: get N points in one interaction.

Session walkthrough

  1. Place the target(s) below the cameras.
  2. Run uv run lens-calibrate.
  3. Press SPACE to measure, or Q to quit.
  4. The Basler rig triangulates all visible tags.
  5. The lens performs a coarse sweep (full diopter range) while the XIMEA camera auto-detects tags and builds per-tag focus ROIs.
  6. For each tag, a fine sweep refines the peak using Gaussian log-space interpolation (Bonatti 2024, eq. 3.9), swept both hi→lo and lo→hi to characterize hysteresis (a warning is printed if the two directions disagree by more than 0.1 D).
  7. Repeat for as many positions as needed.
  8. On quit, a quadratic polynomial D = a·z² + b·z + c is fitted (found to fit as well as the vergence model over the short z ranges involved, and simpler to invert) and a timestamped CSV is written.

Output

Two files are written:

  • calibrations/lens_calib_YYYYMMDD_HHMMSS.csv (relative to the current directory) — full data. Columns: z, diopter, sweep_direction, x, y, n_cameras, n_tags, focus_metric_peak, timestamp. Each measurement contributes two rows per tag — one per sweep direction (hi2lo / lo2hi) — so the fit sees both branches of the hysteresis loop.
  • /home/nfc/src/OptoFly/calibrations/liquid_lens.csv — just z, dpt (the diopter column renamed), which is exactly where OptoFly's liquid_lens.calibration_file expects to find it. If that file already exists, you're prompted replace it with this calibration? [Y/n] — defaults to yes, backing up the old file in place as liquid_lens.csv.bak-YYYYMMDD_HHMMSS before writing the new one. Declining instead saves the new calibration as calibrations/lens_calib_YYYYMMDD_HHMMSS_optofly.csv, alongside the full CSV, and leaves the existing liquid_lens.csv untouched.

To interpolate diopter from z in real time, load the full CSV and either re-fit the polynomial (or the vergence model, if the runtime range grows large enough for its curvature to matter) or use direct interpolation (e.g. numpy.interp).

Algorithm

  • Focus metric: Tenengrad (mean squared Sobel gradient magnitude) — robust against bokeh artifacts on coarse targets like AprilTags.
  • Peak finding: Gaussian log-space 3-point interpolation for sub-step precision; parabola fit and argmax as fallbacks.
  • Coarse sweep: full diopter range, per-tag ROIs derived automatically from AprilTag detections in the XIMEA frame.
  • Fine sweep: ±3 coarse steps around each tag's coarse peak, swept hi→lo then lo→hi (optionally repeated via --fine-repeats) to measure hysteresis.
  • Triangulation: DLT with SVD using the 3×4 world→pixel projection matrices from the braid calibration XML; pixels are undistorted with OpenCV before triangulation.

Module layout

src/liquid_lens_calibration/
├── main.py            CLI loop, CSV output, polynomial fit
├── calibration_io.py  Parse braid XML → per-camera intrinsics + projection matrices
├── cameras.py         Basler camera discovery, frame grab, buffer flush
├── focus_camera.py    XIMEA camera (gain 0, user exposure, full/ROI frame grab)
├── triangulate.py     AprilTag detection, undistortion, DLT triangulation
├── focus.py           Focus metric, coarse+fine sweep, peak interpolation
└── lens.py            Optotune lens wrapper (focal-power mode, diopter sweep)